The Data Logger / Data Logging Instrument has a measuring range of -30 ... 60°C / -22 ... 140°F and can be used in cold environments. In addition to temperature and humidity measurement, the Data Logger / Data Logging Instrument can also measure the absolute air pressure in the ambient atmosphere in a measuring range between 700 ... 1100 hPa. Due to the different sensors, the Data Logger / Data Logging Instrument already covers a wide range of measuring tasks. Typical applications of the Data Logger / Data Logging Instrument are, for example, the monitoring of refrigeration facilities, greenhouses, or an office workplace.
The measured values are stored in the internal memory of the Data Logger / Data Logging Instrument. The Data Logger / Data Logging Instrument has a memory capacity of 13340 measuring points. Once a measurement run has been completed, the stored measurement data can be read out from the Data Logger / Data Logging Instrument independently of the operating system. This is possible because the Data Logger / Data Logging Instrument is recognised as a mass data memory. When storing the measurement data, the measurement points are saved in a .pdf and in a .csv file.
With the corresponding software for the Data Logger / Data Logging Instrument, the measured values can be displayed graphically. In addition, settings for the Data Logger / Data Logging Instrument can be made via the software. For example, the storage interval can be set between 1 minute and 24 hours, and visual alarm limits and start/stop pulses can be set for the Data Logger / Data Logging Instrument. Optionally, the Data Logger / Data Logging Instrument can be equipped with an ISO calibration certificate.
| Absolute pressure | |
| Measuring range | 700 ... 1100 hPa |
| Resolution | 1 hPa |
| Accuracy | ±3 hPa |
| Absolute pressure | |
| Measurement range min. | 525 mmHg |
| Measurement range | 825 mmHg |
| Resolution | 0,1 mmHg |
| Accuracy | ±2.5 mmHg |
| Absolute pressure | |
| Measurement range min. | 20,67 inHg |
| Measurement range | 32,48 inHg |
| Resolution | 0,01 inHg |
| Accuracy | ±0.1 inHg |
| Relative humidity | |
| Measuring range | 0 ... 20 % RH |
| Resolution | 0,1 % RH |
| Accuracy | ±5 % RH |
| Relative humidity | |
| Measuring range | 20 ... 40 % RH |
| Resolution | 0,1 % RH |
| Accuracy | ±3.5 % RH |
| Relative humidity | |
| Measuring range | 40 ... 60 % RH |
| Resolution | 0,1 % RH |
| Accuracy | ±3 % RH |
| Relative humidity | |
| Measuring range | 60 ... 80 % RH |
| Resolution | 0,1 % RH |
| Accuracy | ±3.5 % RH |
| Relative humidity | |
| Measuring range | 80 ... 100 % RH |
| Resolution | 0,1 % RH |
| Accuracy | ±5 % RH |
| Temperature | |
| Measuring range | -30 ... 0 °C |
| Resolution | 0,1 °C |
| Accuracy | ±1.0 °C |
| Temperature | |
| Measuring range | 0 ... 60 °C |
| Resolution | 0,1 °C |
| Accuracy | ±0.5 °C |
| Temperature | |
| Measuring range | -22 ... +32 °F |
| Resolution | 0,1 °F |
| Accuracy | 1.8 °F |
| Temperature | |
| Measuring range | 32 ... 140 °F |
| Resolution | 0,1 °F |
| Accuracy | 0.9 °F |
| General technical data | |
| Display type | LCD |
| Display size | 0,25 Inch |
| Storage medium | Internal memory |
| Storage capacity | 13340 Measurements |
| Storage interval from | 1 min. |
| Storage interval to | 24 h |
| Interface | USB-A |
| Operating time | 3 Year(s) |
| file format | CSV, PDF |
| Menu language | English, English (GB) |
| Protection class (device) | IP20 |
| Power supply | 5 V DC, 0.1 A |
| Connector type | USB-A |
| (Rechargeable) battery | 1 x 3 V CR2450 , lithium |
| Capacity | 600 mAh |
| Operating conditions | -30 ... 60 °C , 0 ... 95 % RH |
| Storage conditions | -30 ... 60 °C , 0 ... 95 % RH |
| Dimensions ( L x W x H ) | 75 x 35 x 15 mm |
| Weight | 33 g |