£ 76.00
| PCE-PQC 34EU | |
| Measuring range | 0.3 ... 25 µm |
| Measuring channel sizes | factory calibrated at 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 µm |
| Counting efficiency | 50% at 0.3 μm, 100% at > 0.45 μm according to JIS |
| Particle mass display | PM0,5, PM1,0, PM2,5, PM5,0, PM10 and TPM |
| Flow rate | 2.83 l/min (0.1 ft³/min) |
| Random loss | 10% at 30000000 particles / ft³ |
| Battery operation | 10 h |
| Light source | long-life laser diode |
| Zero count | <1 count / 5 min (<2 particles/ft³) according to ISO 21501-4 and JIS |
| Counting modes | automatic, manual, real-time display, cumulative/ |
| differential, mass concentration | |
| Acoustic alarm | adjustable, integrated alarm |
| Calibration | traceable to NIST |
| Display | 4.3" WQVGA color touch display, 480x272 px |
| Printer | external thermal printer |
| Air intake | internal pump with automatic |
| flow rate control | |
| Air outlet | interner HEPA filter |
| Battery | replaceable Li-Ion battery |
| Charging time | ca. 4 h |
| Reports | ISO 14644-1 |
| EU GMP appendix 1 | |
| FS 209E | |
| Configuration | memory for 50 user-defined configurations |
| Standards | ISO 21501-4 and JIS B9921 |
| Dimensions | 13.3 x 10.5 x 21 cm |
| Weight | 1.8 kg |
| Memory | 45000 data records (ring memory) consisting of |
| particle counting, temperature and humidity, CO2, places and times | |
| Sample duration | 1 s ... 99 h adjustable |
| Power supply | 110 ... 240 V AC 50/60 Hz |
| Operating conditions | +5 ... +40 °C |
| up to 95% r.H. not condensing | |
| Storage conditions | 0 ... +50 °C |
| up to 98% r.H. not condensing | |
| CO2 Sensor | |
| Measuring range | 0 ... 5000 ppm |
| Resolution | 1 ppm |
| Accuracy | ±1 % of measuring range |
| Temp/humidity sensor internal | 0 ... +50 °C, 15 ... 90 % r.H. |
| Resolution | 0.5 °C |
| Accuracy | ±0.5 °C, ±2 % r.H. |
| Interface | Ethernet, USB |
| Optional interface | Wifi 802.11 b/g, RS485 or RS232 |
| Number of measuring channels | 6 |
| Housing material | stainless steel |