PCE-VC120 Vibration Meter Calibrator
For vibration acceleration, velocity and displacement / Adjustable vibration frequencies
PCE-VC21 is a vibration meter calibrator used to calibrate vibration sensors, meters, analyzers and testers. Suitable for calibrating equipment weighing up to 400 g (0.9 lb), this vibration calibrator generates mechanical vibrations with adjustable quartz-stabilized vibration frequencies between 70 and 10,000 Hz and adjustable vibration amplitudes between 1 m/s² and 20 m/s².
PCE-VC120 features load-independent vibration magnitude with:
- unlimited adjustable vibration frequencies
- less than 2% THD
These adjustable vibration frequencies allow for the calibration of low-frequency vibration sensors such as those in structural vibration meters in line with DIN 4150-3 and in human vibration meters according to ISO 8041.
A control circuit and reference accelerometer inside the excitation or shaker head work in tandem to keep the vibration level constant. A display indicates the error percentage (%). The error % represents the deviation of the target value from the measured amplitude. In settled conditions, the display should come close to 0.0 % if the calibration is correct. If the error % is above 2 %, the device sounds an audible alert to indicate that the error limit is exceeded. In addition, the vibration calibrator switches off automatically in case of overload. Sensors can be connected via magnet or thread. A thread adapter is included in the delivery contents to accommodate M5, M8, 1/4"-28 and UNF 10-32 threads.
The vibration calibrator has many versatile features for field and laboratory use, including a built-in rechargeable battery, battery-saving automatic power-off function, battery charge indicator and AC power adapter for charging. In addition, the supplied carrying case allows for convenient handling and safe transport.
Adjustable vibration frequencies and vibration amplitudes | |||||||||
Frequency (Hz) | 70 to | 10,000 | Hz | ||||||
Max. sensor weight (g), depending on acceleration | |||||||||
at 1 m/s² | 400 | ||||||||
Amplitude accuracy | ± 2 % (80 ... 1000 Hz) ± 5 % (1000 ... 10,000 Hz) | ||||||||
Frequency accuracy | ± 0.05 % | ||||||||
Settling time | < 10 seconds | ||||||||
Level indication | Error percentage (%) display, above ± 3 % audible alert | ||||||||
Max. weight of test object for stated accuracy | 400 g / 0.9 lb | ||||||||
Dynamic force of vibration exciter | 10 N | ||||||||
Max. tightening torque of vibration exciter | 2 Nm | ||||||||
Operating humidity range | < 90 % at 30 °C / 86 °F, non-condensing | ||||||||
Power supply | Built-in NiMH rechargeable battery, 7.2 V / 1.6 Ah | ||||||||
Max. operating hours with rechargeable battery | Approx. 2 ... 4 hours | ||||||||
Charging time | Approx. 4 hours | ||||||||
Charging status | Bar indicator | ||||||||
Automatic power off | 10 minutes / 1 ... 30 minutes adjustable | ||||||||
Stray magnetic field at the vibration head | < 0.2 mT | ||||||||
Dimensions (L x W x H) | Approx. 205 x 105 x 90 mm / 8.1 x 4.1 x 3.5" | ||||||||
Weight | Approx. 3000 g / 106 oz |