Order no.: PCE-PWT 10

Salt Meter PCE-PWT 10

US$ 198.00
:
4250348717399
PCE-PWT 10
ex. VAT, ex. delivery costs
Available in: 1-2 days

Description

Salt Meter PCE-PWT 10

pocket size conductivity tester / for ultrapure water / smallest measuring range 20 µS/cm / salinity measurement / temperature compensation up to 60 °C

The salt meter is a good solution for anyone who wants to check the conductivity of ultrapure water. This salt meter is a compact device with an integrated temperature sensor for optimum temperature compensation. Among other things, the salt meter has a very low measuring range of only 0 ... 20 μS / cm.

Furthermore, the salt meter is able to accurately determine the salinity of liquids. The salt meter has an automatic temperature compensation in the range of 0 ... 60 ° C. The calibration of the salt meter is carried out via a three-point calibration.

Drinking water testing by a salt meter

Water that is usually supplied into the taps for human consumption is transparent and reminds a usual liquid. Though, sometimes, the consumers may notice a strong unpleasant smell, strange metal aftertaste, the change in color or water qualities, when the soap seems to be impossible to wash off the hands and vice versa, when the soap does not lather at all, sometimes water gets a sediment if some amount of it is left still in a glass for a few days or the kettle gets a white limescale just after a few days of exploitation. All these may be the indications that the water has not gone through the correct treatment process and has not been filtered and purified appropriately. Before we get water in the pipes of our households, it undergoes a serious process, during which it should be checked and devoid of any substances that might have negative impact on the health.

Besides such important value as pH (activity of hydrogen ions), that should be checked along with the other parameters, there is also such a notion as the level of TDS, or totally dissolved solids / substances. Water testing by a salt meter in a waste water plant.It does not mean that al the solids that get into water are bad. Some of them have got a natural origin and are nothing but minerals, at the same time, there are lots that end up in our water in the result of many activities that humans are involved into: wastewater, incorrect disposal of waste, industrial processes, discharge in the forbidden places and many others. One of the solutions how to turn water that seems to be unsuitable for consumption because of a high level of various substances present in it is reverse osmosis. The majority of the consumers have a good idea about distilled water, but rarely know about osmosis. These processes not the same, though, similar in that way, that in the result clean water should come out for consumption. Due to the carbon filtration that reverse osmosis applies, water that goes through this process is purified of chlorine. The letter is not that easy to get rid of and the distillers alone do not cope with the task so well as the reverse osmosis devices.

Water after treatment is free from chemicals and harmful substances (chlorine, salts, leads), but there is a belief that it is also free from minerals and necessary elements (magnesium or calcium). Sometimes, water is additionally enhanced with minerals to make it “healthier” to the humans. It is necessary to remember though that high concentration of minerals brings to the increase of pH, it will reduce corrosive effect, but will make water harder. So, when it goes about purification, first thing is to consider is that the drinking water, first of all, and water in general, should have the necessary balance of elements and substances in it, and which is very important – not do harm.

 

One of the stages of water purification is desalting. Separation of salt is a necessary process and can be carried out also with the help of the membranes. The latter are a part of both, electrodialysis and reverse osmosis (it should be fairly noted that, besides these process, there have been many other, but they turned out to be less efficient). Though, the purpose is the same, the principle of these process is different. Reverse osmosis is a successor of electrodialysis and if the latter uses electrical current, the former is based on pressure. For desalting of water also forward osmosis may be applied, which helps to separate the dissolved solutes from the water. There have been long debates as for which type of osmosis is better. Both methods are aimed at improving the water quality, though the operating principles differ, since pretty high pressure if used for Ro, and low for FO. FO has been applied for beverages production.

Due to the development of desalination process, even sea water may be turned into potable, though, of course, it is a more cost and energy consuming process if to compare with fresh water, where it goes about purification.

Ability to purify water has brought an incredible contribution into various spheres of life, not only technology and industry. It significantly improves many processes, helps to preserve equipment in a better condition (without excessive corrosions and limescale), it helps in agriculture and irrigation, and of course, it provides people with pure and healthy drinking water.


Highlights

- smallest measuring range up to 20 µS/cm
- compact design
- automatic temperature compensation
- salinity measurement
- internal three-point calibration
- changeable electrode

Specification

Conductivity measuring range 0 ... 20 µS/cm
0 ... 200 µS/cm
0 ... 2000 µS/cm
0 ... 20.00 mS/cm
Resolution 0.01 / 0.1 / 1
Accuracy ± 1 % of measuring range
   
Salinity measuring range 0.00 ... 10.00 ppt
Resolution 0.01 ppt
Accuracy ± 1 % of measuring range
   
Temperature measuring range  0 ... 60 °C / 32 ... 140 °F
Resolution 0.1 °C
Accuracy ± 1 °C
   
Temperature compensation range 0 ... 60 °C / 32 ... 140 °F
Temperature coefficient 2 % / °C
Number of calibration points 1 ... 3
Calibration points 84 µS/cm
1413 µS/cm
12.88 mS/cm
Cell constant K=1
Normal media temperature 25 °C / 77 °F
Hold function By pressing a button
Automatic shutdown 8 minutes after the last key press
Pause Reset to factory setting possible
Power supply 2 x 1.5 V AAA batteries
Operating time ca. 200 hours
Dimensions 185 x 40 mm / 7.2 x 1.5"
Weight ca. 100 g / 3.5 oz

Delivery Scope

1 x Salt Meter PCE-PWT 10
2 x 1.5 V AAA batteries
1 x User manual