The Current Clamp has a measuring range of 0… 1000 A AC/DC. The current is measured inductively using the clamp on the measuring device. Live supply lines with a diameter of up to 32 mm (1.2") can be connected to the Current Clamp. With the optional Rogowski coil, the measuring range of the Current Clamp can be extended to up to 3000 A AC. The Current Clamp has sub-functions for current measurement. The inrush current can be measured with the Current Clamp. This is a particularly important function, as motors such as those installed in ventilation systems or hall lighting require a particularly high current when switched on.
In addition to current measurement, the Current Clamp can also measure voltages from 0… 1000 V AC/DC, resistances, frequencies and capacitances. The functions can be set using the rotary selector switch on the Current Clamp. With the NCV function, the Current Clamp can be used to determine the absence of voltage without contact. As soon as a voltage of 150 V AC is applied, this is indicated by an additional red LED. This function is active for every measuring function selected using the rotary selector switch. An LED lamp is integrated in the Current Clamp for optimal illumination of the measuring point.
The Current Clamp can be connected to an Android or iOS device via the Bluetooth interface. With the free app for the Current Clamp, the measured values can be displayed, saved and further processed. The Current Clamp thus covers a wide range of measuring tasks due to the multitude of functions.
| Direct voltage DC | |
| Measuring range | 0 ... 600 mV |
| Resolution | 0,1 mV |
| Accuracy | ±(0.5 % +8 digits) |
| Overload | 1000 V AC/DC |
| Direct voltage DC | |
| Measuring range | 0 ... 6 V |
| Resolution | 0,001 V |
| Accuracy | ±(1.5 % +5 digits) |
| Direct voltage DC | |
| Measuring range | 0 ... 60 V |
| Resolution | 0,01 V |
| Accuracy | ±(1.5 % +5 digits) |
| Direct voltage DC | |
| Measuring range | 0 ... 600 V |
| Resolution | 0,1 V |
| Accuracy | ±(1.5 % +5 digits) |
| Direct voltage DC | |
| Measuring range | 0 ... 1000 V |
| Resolution | 1 V |
| Accuracy | ±(1.5 % + 5 digits) |
| Direct current DC | |
| Measuring range | 0 ... 600 A |
| Resolution | 0,1 A |
| Accuracy | ±(2.5 % +5 digits) |
| Direct current DC | |
| Measuring range | 0 ... 1000 A |
| Resolution | 1 A |
| Accuracy | ±(2.8% +5 digits) |
| Capacity | |
| Measuring range | 0 ... 60 nF |
| Resolution | 0,01 nF |
| Accuracy | ±(4% +20 digits) |
| Capacity | |
| Measuring range | 0 ... 600 nF |
| Resolution | 0,1 nF |
| Accuracy | ±(3% +8 digits) |
| Capacity | |
| Measuring range | 0 ... 6 µF |
| Resolution | 0,001 µF |
| Accuracy | ±(3% +8 digits) |
| Capacity | |
| Measuring range | 0 ... 60 µF |
| Resolution | 0,01 µF |
| Accuracy | ±(3% +8 digits) |
| Capacity | |
| Measuring range | 0 ... 600 µF |
| Resolution | 0,1 µF |
| Accuracy | ±(3% +8 digits) |
| Capacity | |
| Measuring range | 0 ... 6000 µF |
| Resolution | 1 µF |
| Accuracy | ±(5% +8 digits) |
| Capacity | |
| Measuring range | 0 ... 60 mF |
| Resolution | 0,01 mF |
| Accuracy | ±(5% +20 digits) |
| Capacity | |
| Measuring range | 0 ... 100 mF |
| Resolution | 0,1 mF |
| Accuracy | ±(5% +20 digits) |
| Alternating current AC | |
| Measuring range | 0 ... 600 A |
| Resolution | 0,1 A |
| Accuracy | ±(2.5 % +5 digits) |
| Overload | 1000 A AC |
| Frequency range | 50 ... 60 Hz |
| Alternating current AC | |
| Measuring range | 0 ... 1000 A |
| Resolution | 1 A |
| Accuracy | ±(2.8% +5 digits) |
| Frequency range | 50 ... 60 Hz |
| Resistance | |
| Measuring range | 0 ... 600 Ω |
| Resolution | 0,1 Ω |
| Accuracy | ±(1% +10 Digits) |
| Resistance | |
| Measuring range | 0 ... 6 kΩ |
| Resolution | 0,001 kΩ |
| Accuracy | ±(1.5 % +5 digits) |
| Resistance | |
| Measuring range | 0 ... 60 kΩ |
| Resolution | 0,01 kΩ |
| Accuracy | ±(1.5 % +5 digits) |
| Resistance | |
| Measuring range | 0 ... 600 kΩ |
| Resolution | 0,1 kΩ |
| Accuracy | ±(1.5 % +5 digits) |
| Resistance | |
| Measuring range | 0 ... 6 M Ω |
| Resolution | 0,001 M Ω |
| Accuracy | ±(2.5 % +5 digits) |
| Resistance | |
| Measuring range | 0 ... 60 M Ω |
| Resolution | 0,01 M Ω |
| Accuracy | ±(3.5% +10 Digits) |
| Frequency | |
| Measuring range | 9,999 Hz ... 99,9 kHz |
| Resolution | 10 Hz |
| Accuracy | ±(1.2 % +5 digits) |
| Temperature | |
| Measuring range | -40 ... +600 °C |
| Resolution | 0,1 °C |
| Accuracy | ±(1.5 % +5 digits) |
| Temperature | |
| Measuring range | 600 ... 1000 °C |
| Resolution | 1 °C |
| Accuracy | ±(1.5 % +5 digits) |
| Temperature | |
| Measuring range | -40 ... +1112 °F |
| Resolution | 0,1 °F |
| Accuracy | ±(1.5% +9 Digits) |
| Temperature | |
| Measuring range | 1112 ... 1800 °F |
| Resolution | 1 °F |
| Accuracy | ±(1.5% +9 Digits) |
| General technical data | |
| Display type | LC colour display |
| Measuring rate | 3 x per second |
| Interface | Wireless |
| Maximum cable diameter | 34 mm |
| Automatic power-off from...to | 15 ... 60 min. |
| Dimensions | Clamp opening: 34 mm / 1.3 inches |
| Diode test | Test current: <1.5 mA Maximum voltage with open circuit: 3.3V DC |
| Menu language | English, English (GB) |
| Protection class (device) | IP20 |
| Power supply | batteries |
| (Rechargeable) battery | 3 x 1,5 V AAA battery , Alkali-manganese |
| Capacity | 1200 mAh |
| Operating conditions | 5 ... 40 °C , 0 ... 80 % RH |
| Storage conditions | -20 ... 40 °C , 0 ... 80 % RH |
| Dimensions ( L x W x H ) | 250 x 80 x 40 mm |
| Weight | 210 g |